Skip to content
Recall Observatory FDA recall evidence

Device product

Description/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYSIS KIT 2L: 14 FR X 20CM AGB/AK-25142-CDC; HEMODIALYSIS KIT: 2L 14 FR X 15 CM/ASK-22142-UMD; HEMODIALYSIS KIT: 2L 14 FR X 15 CM/ASK-22142-UNC; HEMODIALYSIS KIT: 2L 14 FR X 20 CM/ASK-25142-UMD; HEMODIALYSIS KIT 2L: 14 FR X 20 CM AGB/ASK-25142-UN1; HEMODIALYSIS KIT: 2L 14 FR X 25 CM/ASK-26142-UMD; HEMODIALYSIS KIT: 2LUMEN 12 FR X 13 CM/AU-23122-F; HEMODIALYSIS KIT: 2LUMEN 12 FR X 20 CM/AU-25122-F; HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-X1A; HEMODIALYSIS KIT: 2L 12 FR X 16 CM AGB/CDC-22122-XU1A; HEMODIALYSIS KIT: 2L 14 FR X 15 CM AGB/CDC-22142-X1A; HEMODIALYSIS KIT: 2L 12 FR X 25 CM AGB/CDC-26122-X1A; HEMODIALYSIS KIT 2L: 12 FR X 16CM AGB/AK-22122-CDC; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122-F; HEMODIALYSIS KIT: 2-LUMEN 14 FR X 15 CM/AK-22142-F; AHDC AGB U-BEND KIT: 2-L 12 FR X 16 CM/CDC-22122-CN1AU; AHDC AGB CURVED KIT: 2-L 14 FR X 15 CM/CDC-22142-CN1AC; AHDC AGB U-BEND KIT: 2-L 12 FR X 13 CM/CDC-23122-CN1AU; AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-CN1AU; AHDC AGB CURVED KIT: 2-L 14 FR X 20 CM/CDC-25142-CN1AC;

Z-2373-2026

April 21, 2026

Class I

Product summary

Firm
ARROW INTERNATIONAL, LLC
Event
Event 98793
Status
Ongoing
Classification
Class I
Quantity
15,425
Official record key
device-enforcement:Z-2373-2026

Official wording

Reason: Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Code information: REF(Material)/UDI-DI/Lot(Batch): AK-11802-F/10801902193275, 10801902197129/33F24G0344, 33F24J0511, 33F24L0637, 33F25A0473, 33F25E0060, 33F25J0473, 33F24A0848; AK-22122/10801902193275/33F24H0232, 33F24J0523; AK-22142-CDC/10801902193275/33F24J0527, 33F24K0129, 33F24L0649, 33F25C0115, 33F25C0959, 33F25E0940, 33F25J0481, 33F25L0428, 33F24G0364; AK-25142-CDC/10801902193275, 50801902123195/33F25A0486, 33F25C0892, 33F25E0086, 33F25E0944, 33F25J0485, 33F25J0845, 33F25L0374, 33F26B0521, 33F25A0487; ASK-22142-UMD/10801902193275, 10801902136470, 10801902143690/33F25F0362, 33F25J0094, 33F26A0244, 33F26B0137, 33F24F0041; ASK-22142-UNC/10801902193275/33F25G0269, 33F24K0560; ASK-25142-UMD/10801902193275, 10801902136500/33F25B0239, 33F25E0308, 33F25G0686, 33F25K0425, 33F26A0246, 33F26C0047, 33F24K0696; ASK-25142-UN1/10801902193275, 10801902156379/33F24L1157, 33F26B0908, 33F25B0420; ASK-26142-UMD/10801902193275, 33F26A0408, 33F25B0427; AU-23122-F/10801902193275/33F25C0877, 33F25H0157, 33F24L0038; AU-25122-F/10801902193275, 70801902123328, 10801902210071/33F25G0562, 33R24L0038, 33F24D0400; CDC-22122-X1A/10801902193275, 10801902213478/33F25E0694, 33F25G0268, 33F24D0693; CDC-22122-XU1A/10801902193275, 10801902168044/33F25E0652, 33F25H0643, 33F24D0830; CDC-22142-X1A/10801902193275/33F25A0284; CDC-26122-X1A/10801902213515/33F24F1028; AK-22122-CDC/10801902193275, 10801902198492/33F24K0125, 33F24L0647, 33F25C0114, 33F25E0936, 33F25G0091, 33F24E0355; AK-22122-F/10801902193275/33F24G0360, 33F24J0524, 33F24K0126, 33F25A0482, 33F25E0075, 33F25E0937, 33F25J0839, 33F25L0413, 33F24G0361; AK-22142-F/10801902193275/33F24L0662, 33F25C0117, 33F25C0865, 33F25H0034, 33F25J0841, 33F24K0130; CDC-22122-CN1AU/10801902193275, 10801902222265/33F26A1388, 33F26B0398, 33F24L0966; CDC-22142-CN1AC/10801902193275/33F26A1550, 33F25J0455; CDC-23122-CN1AU/10801902193275, 10801902222319/33F26A0008, 33F26A1478, 33F26B0397, 33F24L0279; CDC-25122-CN1AU/10801902193275/33F25M0346, 33F25E0882; CDC-25142-CN1AC/30801902123597/36F25B0018

Distribution pattern: US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.