Device product
Philips Azurion 3M12, Model Numbers: 722063, 722221 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Z-2284-2026
Product summary
- Event
- Event 98977
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 16 units
- Official record key
device-enforcement:Z-2284-2026
Official wording
Reason: During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Code information: UDI (01)00884838099203(21)3, (01)00884838099203(21)7, (01)00884838085275(21)131, (01)00884838099203(21)54, (01)00884838085275(21)84, (01)00884838099203(21)6, (01)00884838099203(21)93, (01)00884838085275(21)132, (01)00884838099203(21)69, (01)00884838099203(21)9, (01)00884838099203(21)115, (01)00884838099203(21)85, (01)00884838099203(21)19, (01)00884838099203(21)49, (01)00884838099203(21)108, (01)00884838085275(21)108.
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.