Skip to content
Recall Observatory FDA recall evidence

Device product

Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.

Z-2262-2026

April 22, 2026

Class II

Product summary

Firm
Baxter Healthcare Corporation
Event
Event 98852
Status
Ongoing
Classification
Class II
Quantity
105675
Official record key
device-enforcement:Z-2262-2026

Official wording

Reason: Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.

Code information: Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806

Distribution pattern: Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.