Philips Azurion 5M12, Model Numbers: 722227 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
Z-2286-2026
Product summary
- Event
- Event 98977
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 3 units
- Official record key
device-enforcement:Z-2286-2026
Official wording
Reason: During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Code information: UDI (01)00884838099227(21)88, (01)00884838099227(21)135, (01)00884838099227(21)63.
Distribution pattern: US Nationwide distribution.
Derived failure modes
-
Unknown
During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.