Device product
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Z-2243-2026
Product summary
- Event
- Event 98938
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 14 units
- Official record key
device-enforcement:Z-2243-2026
Official wording
Reason: GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Code information: UDI none, Serial number or Sales Order number SS921120014SA, SS921120017SA, SS921120021SA, SS921120018SA, SS921120020SA, SS920390001SA, SS921120007SA, SS921120008SA, SS921120012SA, SS921120016SA, SS920500001SA, SS920500002SA, SS921120009SA, SS921120010SA.
Distribution pattern: Worldwide distribution.
Derived failure modes
-
Unknown
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).