Device product
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Z-1980-2026
Product summary
- Event
- Event 98666
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 18,080 units
- Official record key
device-enforcement:Z-1980-2026
Official wording
Reason: Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Code information: Material Number: 10872079. Model Number: DF33B. UDI: (01)00630414595009(10)GA6307(17)20261103, (01)00630414595009(10)BA7005(17)20270105. Lot Numbers: GA6307, BA7005.
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.
Derived failure modes
-
Unknown
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results