Device product
Soft-Vu Angiographic Catheter, Straight Art, Non-Braided; SOFT-VU STR ART 4F X 65CM 035 NB 0SH; Catalog No.: 10731401; Product/UPN No.: H787107314015 (Box), H787107314010 (Pouch); Box Quantity: 5 pouches;
Z-1869-2026
Product summary
- Event
- Event 98463
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 220 units (44 boxes)
- Official record key
device-enforcement:Z-1869-2026
Official wording
Reason: The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.
Code information: SOFT-VU STR ART 4F X 65CM 035 NB 0SH; Catalog No.: 10731401; Product/UPN No.: H787107314015 (Box), H787107314010 (Pouch); UDI-DI: 25051684009475(Box), 15051684009478(Pouch); Lot No.: A2825042;
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, China, Chile, Curacao, Czech Republic, Germany, Ireland, Israel, Italy, Lebanon, Netherlands, New Zealand, Panama, Slovenia, South Africa, Spain, Sweden, Thailand.
Derived failure modes
-
Manufacturing or process control
manufacturing defect