Device product

Aurous Centimeter Sizing Catheter RPN N5.0-35-100-P-10S-PIG-CSC-20-01 GPN G47308 Sizing catheters are single-use, sterile catheters with gold marker bands. These catheters are for use in angiographic procedures and available in a variety of lengths (70 and 100 cm) and tip configurations (straight, PIG, and VCF).

Z-1935-2026

April 02, 2026

Class I

Product summary

Firm
Cook Incorporated
Event
Event 98682
Status
Ongoing
Classification
Class I
Quantity
271
Official record key
device-enforcement:Z-1935-2026

Official wording

Reason: Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.

Code information: UDI-DI 00827002473089 Lots 16858459 16869145 16857138 16763324 16750208 16752733 16821053 16713191

Distribution pattern: US distribution nationwide. OUS distribution to Brunei, Brazil, Canda, China, Colombia, Costa Rica, Hong Kong, Israel, India, Japan, South Korea, Macau, New Zealand, Singapore, El Salvador, Taiwan, Uruguay

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Marker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario, device fragmentation and separation may cause life-threatening harm or death.