Skip to content
Recall Observatory FDA recall evidence

Device product

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PACK, Medline Kit Number/SKU DYNJ62635B.

Z-2018-2026

March 19, 2026

Class II

Product summary

Firm
Medline Industries, LP
Event
Event 98664
Status
Ongoing
Classification
Class II
Quantity
208 kits
Official record key
device-enforcement:Z-2018-2026

Official wording

Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Code information: Medline Kit Number/SKU DYNJ32854B: UDI/DI each 10198459413414, UDI/DI case 40198459413415, Lot Number : 25GBU656; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25GMD214; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25FMC376; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25DMJ030; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25BME902; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 25AMJ032; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24IMH612; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24EME486; Medline Kit Number/SKU DYNJ62635B: UDI/DI each 10195327634780, UDI/DI case 40195327634781, Lot Number : 24DMD315; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 26BMD607; Medline Kit Number/SKU DYNJ62635C: UDI/DI each 10198459485619, UDI/DI case 40198459485610, Lot Number : 25JMD196.

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Potency or specification failure

    reason.potency_specification_failure · v1.0.0

    out-of-specification