Skip to content
Recall Observatory FDA recall evidence

Device product

B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG DR. Model Number: SL-2010M2096A.

Z-1798-2026

March 19, 2026

Class I

Product summary

Firm
B Braun Medical Inc
Event
Event 98552
Status
Ongoing
Classification
Class I
Quantity
3,158,104 units
Official record key
device-enforcement:Z-1798-2026

Official wording

Reason: Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Code information: Model Number SL-2010M2096A; UDI-DI Primary: 04046955674992; UDI-DI Unit: 04046955735280; All lots manufactured since 09JUN2025

Distribution pattern: Worldwide distribution - US Nationwide and the country of Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.