Device product
Medline medical convenience kits, containing NAMIC Angiographic Rotating Adaptor (RA) Control Syringes (Namic RA syringes), labeled as follows: SPECIAL PROCEDURE TRAY, Medline SKU # DYNJ80513B
Z-1720-2026
Product summary
- Event
- Event 98598
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 762 units
- Official record key
device-enforcement:Z-1720-2026
Official wording
Reason: Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.
Code information: Medline SKU # DYNJ80513B, UDI/DI each 10198459389375, UDI/DI case 40198459389376, Lot Numbers: 26ABH752, 25LBL571, 25KBO421, 25IBF501, 25HBL993, 25FBU283.
Distribution pattern: Worldwide distribution: US (nationwide) including Puerto Rico (PR) and OUS (foreign) to countries of: Canada (CA), Netherlands (NL), Australia (AU), Korea, Republic of (KR), Sri Lanka (LK), Pakistan (PK), Japan (JP), United Arab Emirates (AE), Singapore (SG) and Slovakia (SK).
Derived failure modes
-
Unknown
Medline Industries, LP has identified through post-market surveillance a potential risk of the syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.