Device product
FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.
Z-1605-2026
Product summary
- Firm
- I.T.S. GmbH
- Event
- Event 98341
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 15,517 units
- Official record key
device-enforcement:Z-1605-2026
Official wording
Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Code information: 1. Article Number (Lot Numbers): 37303-10 (18/151720, 18/268729, 679/032432, 679/072268, 679/102237, 679/102307), 37303-12 (18/570637, 679/032434, 679/072269, 679/102346, 679/112227), 37303-14 (18/751091, 26/F66913, 679/012510, 679/062288, 679/102359, 679/112410), 37303-16 (679/072263, 679/102245, 18/151841, 26/F66914, 18/331193, 18/543625, 679/012511, 679/042403, 679/102360), 37303-18 (18/151942, 18/210467, 18/231415, 18/341194, 18/572633, 18/771093, 679/012512, 679/022507, 679/042404, 679/102248, 679/112319), 37303-20 (18/152043, 18/211468, 18/351195, 18/573634, 18/701585, 18/781094, 679/022508, 679/042420, 679/112216, 679/112328), 37303-22 (18/171705, 18/212469, 18/361196, 18/574635, 18/791095, 679/022509, 679/042417, 679/112224, 679/112340), 37303-24 (18/575639, 18/741572, 26/F66915, 679/022510, 679/112226, 679/122313, 679/122333), 37303-26 (18/152346, 18/214471, 26/F66916, 679/012403, 679/112230, 679/122323), 37303-28 (26/F66917, 679/112231, 679/122328), 37303-30 (18/192716, 18/591743, 679/122334). 2. Article Number (Lot Numbers): 32271-10 (18/061949, 18/231167, 679/062351), 32271-12 (18/048927, 26/F66235, 679/042215, 679/062342, 679/062405, 679/062489, 679/072271), 32271-14 (18/049928, 18/231165, 26/F66140, 679/042210, 679/052340, 679/052518, 679/062471, 679/062490, 679/072267), 32271-16 (18/231164, 18/311125, 26/F66141, 679/042216, 679/052519, 679/062315, 679/062483, 679/072272, 679/072401), 32271-18 (18/062950, 26/F65556, 26/F66236, 679/032261, 679/062332, 679/072402), 32271-20 (18/063951, 26/F66237, 679/052326), 32271-22 (18/064952, 26/F65557, 679/032263, 679/062354), 32271-24 (18/065953, 26/F65558, 679/062333), 32271-26 (679/032265), 32271-28 (26/F65559, 679/062357, 679/052520), 32271-32 (26/F65561), 32271-34 (26/F65562). 3. Article Number (Lot Numbers): 37304-10 (18/635486, 26/F24799, 679/012419, 86/011176), 37304-12 (18/061352, 18/521172, 18/5211721, 18/576643, 18/630487, 679/012433, 679/092207, 679/122409), 37304-14 (18/081360, 18/531173, 18/5311731, 18/577644, 18/631488, 679/012434, 679/092208, 679/122410), 37304-16 (18/541174, 18/578645, 18/632489, 679/012444, 679/092201, 679/122411), 37304-18 (18/092020, 18/551175, 18/633490, 679/022403, 679/092202), 37304-20 (18/011320, 18/561176, 18/634491, 679/012287, 679/012420, 679/082217, 86/483331), 37304-22 (18/241273, 679/012288, 679/012421), 37304-24 (18/021321, 18/571177, 679/012289, 679/012445). 4. Article Number (Lot Numbers): 21413-6 (26/F52124). 5. Article Number (Lot Numbers): 21431-ME (679/102045). 6. Article Number (Lot Numbers): 21482 (679/092020), 21483 (26/F55902). 7. Article Number (Lot Numbers): 21421-5 (26/F65377, 679/092039). 8. Article Number (Lot Numbers): 37241-10 (18/012169, 18/610104, 18/701013), 37241-12 (18/231295, 18/767108), 37241-14 (18/511016, 18/537392, 26/F24807, 679/012280), 37241-16 (18/005340, 18/121106, 18/231387, 18/521017, 18/560517), 37241-18 (18/232328, 18/531018, 679/072408, 86/311848), 37241-20 (18/233388, 18/541020, 18/562519, 679/072419, 86/287847), 37241-22 (18/066109, 18/234389, 18/551021, 26/F24594, 26/F24811), 37241-24 (18/561022, 18/564521, 18/608110, 26/F24595), 37241-26 (18/611111, 18/711023, 26/F24596, 26/F24813), 37241-28 (18/612112, 18/721024, 18/811597, 26/F24597, 86/485508), 37241-30 (18/015172, 18/613113, 18/731019, 26/F24598). 9. Article Number (Lot Numbers): 21015-08 (00/0123-12, 00/0321-43, 00/0723-07), 21015-09 (00/0123-13).
Distribution pattern: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Derived failure modes
-
Unknown
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).