Device product
I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.
Z-1598-2026
Product summary
- Firm
- I.T.S. GmbH
- Event
- Event 98341
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 510 units
- Official record key
device-enforcement:Z-1598-2026
Official wording
Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Code information: 1. Article Number (Lot Numbers): 21604-6 (643/052301, 643/0652, 643/092201), 21604-9 (643/032311, 643/062308, 643/062309, 643/112310, 643/122201). 2. Article Number (Lot Numbers): 21605-11 (643/072314, 643/102207), 21605-6 (643/0654, 643/072306, 643/072316, 643/072338, 643/102208, 643/102212, 643/102217, 643/112306, 643/122202), 21605-9 (643/0643, 643/0662, 643/072307, 643/072339, 643/112309, 643/122203). 3. Article Number (Lot Numbers): 21105-6 (86/876787), 21105-8 (86/876788). 4. Article Number (Lot Numbers): 21104-4 (679/052234, 679/082305, 86/036334), 21104-5 (679/0722101, 679/082306, 86/036335), 21104-6 (679/012307, 86/876760), 21104-7 (679/082307, 86/036336, 86/876781), 21104-8 (679/082308, 86/876783), 21104-9 (86/876904). 5. Article Number (Lot Numbers): 21101-5 (86/876637), 21101-6 (679/121605, 00/1105-12, 26/F41923, 86/876638, 86/903610), 21101-7 (770/201905, 00/0518-06, 679/042303, 86/876640), 21101-8 (86/876721, 679/062224), 21101-9 (679/122264, 86/876727), 21101-11 (86/876633), 21101-13 (86/876635). 6. Article Number (Lot Numbers): 21102-6 (770/201905). 7. Article Number (Lot Numbers): 21103-8 (679/092216), 21103-9 (679/062215, 86/876758), 21103-10 (679/092219, 86/876741), 21103-11 (679/062214), 21103-12 (86/876743), 21103-13 (679/022303, 86/876745).
Distribution pattern: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Derived failure modes
-
Unknown
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).