Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;

Z-1633-2026

March 17, 2026

Class II

Product summary

Firm
Straumann USA LLC
Event
Event 98565
Status
Ongoing
Classification
Class II
Quantity
7 units
Official record key
device-enforcement:Z-1633-2026

Official wording

Reason: A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Code information: Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;

Distribution pattern: Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.