Device product
da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45
Z-1355-2026
Product summary
- Event
- Event 98205
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 47
- Official record key
device-enforcement:Z-1355-2026
Official wording
Reason: Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.
Code information: UDI: 00886874119747, Serial: SQ0326, SQ0336, SQ0341, SQ0345, SQ0346, SQ0356, SQ0367, SQ0368, SQ0375, SQ0377, SQ0379, SQ0380, SQ0386, SQ0387, SQ0389, SQ0394, SQ0395, SQ0398, SQ0401, SQ0404, SQ0412, SQ0413, SQ0415, SQ0419, SQ0420, SQ0422, SQ0425, SQ0427, SQ0428, SQ0432, SQ0433, SQ0436, SQ0443, SQ0447, SQ0448, SQ0451, SQ0454, SQ0457, SQ0458, SQ0462, SQ0464, SQ0469, SQ0470, SQ0662, SQ0727, SQ0740, SQ0761
Distribution pattern: US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.
Derived failure modes
-
Device software or design
firmware, may result in