Skip to content
Recall Observatory FDA recall evidence

Device product

Olympus High Flow Insufflation Unit. Model Number: UHI-3. Used for insufflation of the abdominal cavity, and automatic suction and smoke evacuation to facilitate laparoscopic observation and treatment within the abdominal cavity.

Z-1354-2026

January 16, 2026

Class I

Product summary

Firm
Olympus Corporation of the Americas
Event
Event 98243
Status
Ongoing
Classification
Class I
Quantity
485 units
Official record key
device-enforcement:Z-1354-2026

Official wording

Reason: Issue with software algorithm which may lead to overpressure events.

Code information: Model Number: UHI-3; UDI-DI: 04953170140280, 04953170140297; All Serial Numbers.

Distribution pattern: Nationwide distribution

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    Issue with software