Skip to content
Recall Observatory FDA recall evidence

Device product

HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 6mm x 8mm; Material Number (UPN): M00553520;

Z-1161-2026

December 19, 2025

Class I

Product summary

Firm
Boston Scientific Corporation
Event
Event 98237
Status
Ongoing
Classification
Class I
Quantity
266 units
Official record key
device-enforcement:Z-1161-2026

Official wording

Reason: Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

Code information: Material Number (UPN): M00553520; UDI-DI: 8714729904540; Lot Numbers: 36067957, 36067958, 36067959, 36067960, 36067961, 36099410, 36102139, 36111317, 36111318, 36111319, 36132439, 36132800, 36136724, 36136725, 36144054, 36154527, 36154528, 36154529, 36156019, 36164532, 36164990, 36176206, 36176207, 36177448, 36191239, 36199389, 36199390, 36217761, 36241713, 36251361, 36252913, 36271441, 36271442, 36281910, 36281911, 36292639, 36292680, 36292681, 36293836, 36300838, 36317858, 36317859, 36318240, 36323786, 36323787, 36331036, 36331037, 36331038, 36331039, 36397921, 36397922, 36397923, 36421238, 36424644, 36424645, 36427248, 36446498, 36447200, 36447201, 36449569, 36449570, 36449571, 36458829, 36459567, 36459568, 36459569, 36459570, 36461959, 36462320, 36468392, 36468393, 36468394, 36502795, 36506161, 36506163, 36509326, 36520053, 36523274, 36528114, 36536932, 36536933, 36548030, 36548031, 36548850, 36548851, 36552680, 36559035, 36559036, 36567446, 36567447, 36567448, 36574150, 36581585, 36581586, 36582790, 36582791, 36585188, 36585189, 36591087, 36592499, 36592660, 36596040, 36596041, 36609818, 36609819, 36610400, 36610401, 36621583, 36621584, 36634546, 36634547, 36664793, 36673195, 36673557, 36673558, 36675688, 36675689, 36675690, 36675692, 36677844, 36677845, 36677846, 36680231, 36680232, 36683260, 36683261, 36683262, 36693971, 36693972, 36717713, 36717714, 36717715, 36717819, 36721863, 36731805, 36731806, 36738965, 36738966, 36743835, 36750922, 36750924, 36753179, 36753400, 36781577, 36781578, 36781579, 36784378, 36784379, 36788332, 36788333, 36788334, 36792650, 36799316, 36832750, 36832837, 36832838, 36833682, 36834573, 36843542, 36844231, 36846161, 36846163, 36864951, 36864952, 36865501, 36865502, 36865504, 36865553, 36868065, 36872344, 36872346, 36878419, 36884609, 36884612, 36884613, 36895367, 36895368, 36896664, 36907033, 36907249, 36907378, 36907400, 36914565, 36915206, 36915207, 36916329, 36916330, 36917403, 36917404, 36917993, 37000847, 37003365, 37423232, 37423233, 37427258, 37427259, 37427360, 37427361, 37427362, 37437936, 37438283, 37438286, 37438287, 37439978, 37440320, 37451304, 37456594, 37457216, 37457217, 37457269, 37459571, 37463623, 37463624, 37463625, 37463628, 37469756, 37469757, 37469758, 37470669, 37472148, 37475121, 37475122, 37481983, 37486079, 37518915, 37518916, 37518917, 37519094, 37532420, 37532421, 37533309, 37543372, 37558499, 37558542, 37567252, 37567253, 37567400, 37568848, 37568849, 37571930, 37685005, 37686597, 37689957, 37689958, 37698144, 37704842, 37761033, 37761034, 37775246, 37776579, 37776720, 37776721, 37780826, 37785688, 37785689, 37811064, 37817121, 37823587, 37823588, 37826529, 37826530, 37827570, 37834839, 37836866, 37836867, 37842827;

Distribution pattern: Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.