Skip to content
Recall Observatory FDA recall evidence

Device product

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

Z-1308-2026

November 17, 2025

Class II

Product summary

Firm
ETAC A/S
Event
Event 98087
Status
Ongoing
Classification
Class II
Quantity
95 units
Official record key
device-enforcement:Z-1308-2026

Official wording

Reason: A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

Code information: (1) Model/Catalog Number: M14000; UDI-DI: 05707995032270; Serial Numbers: 616606, 616604, 591003, 591002, 590906, 590901, 590904, 590903, 590902, 591001, 604101, 604005, 591703, 591007, 591006, 591005, 591004, 615105, 615104, 615103, 615102, 615101, 604004, 604003, 604002, 604001, 614202, 614201, 591702, 591701, 616603, 615106, 590905; (2) Model/Catalog Number: M1401; UDI-DI: 05707995032270; Serial Numbers: 593302, 593301, 593303, 612507, 612506, 612508, 612505, 612503, 620702, 612504, 612502, 612501, 593304, 593305, 620703, 620701; (3) Model/Catalog Number: M1402; UDI-DI: 05707995032270; Serial Numbers: 612601; (4) Model/Catalog Number: M14050; UDI-DI: 05707995032270; Serial Numbers: 590801, 596701, 585701, 584501, 591801, 604405, 604406, 604402, 604404, 604403, 604401, 604303, 604302, 604301; (5) Model/Catalog Number: M14150; UDI-DI: 05707995032270; Serial Numbers: 585811, 585810, 585809, 585808, 585806, 585805, 585804, 585803, 585801, 596604, 596603, 596602, 596601, 596611, 596610, 596609, 596608, 596607, 596606, 596605, 585807, 585802; (6) Model/Catalog Number: M1402-01; UDI-DI: 05707995032270; Serial Numbers: 612701, 612703, 612702; (7) Model/Catalog Number: M14050-00; UDI-DI: 05707995032270; Serial Numbers: 584307; (8) Model/Catalog Number: M14001; UDI-DI: 05707995032270; Serial Numbers: 603501, 599801, 599802, 604202, 604201;

Distribution pattern: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.