Skip to content
Recall Observatory FDA recall evidence

Device product

Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554¿

Z-1323-2026

December 22, 2025

Class I

Product summary

Firm
Medline Industries, LP
Event
Event 98277
Status
Ongoing
Classification
Class I
Quantity
650
Official record key
device-enforcement:Z-1323-2026

Official wording

Reason: These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025

Code information: UDI-DI (case) 20197344019554¿ (ea) 10197344019557 Lots EP250324 EP250416 EP250512 EP250521 EP250609 EP250630 EP250710 EP250724

Distribution pattern: Distribution US nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025