Device product
Brand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 16.0.0 Software Version: RayStation 2024B Product Description: Radiation Therapy Treatment Planning System Component: No
Z-1109-2026
Product summary
- Event
- Event 98117
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 0
- Official record key
device-enforcement:Z-1109-2026
Official wording
Reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Code information: Lot Code: UDI: 0735000201077820240625 GTIN: 07350002010778 Serial Numbers: 16.0.0.847 Software Revision: RayStation 2024B Expiration Date: 2029-06-28
Distribution pattern: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Derived failure modes
-
Unknown
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.