Device product
Conductivity Standard Solution Catalog/Model Numbers: 10001, 10011, 10021, 10032, 10040, 10040c, and 10041 Conductivity Standard Solutions are a secondary standard solution used for the calibration of conductivity cells together with conductivity meters
Z-1185-2026
Product summary
- Event
- Event 98030
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 499 units
- Official record key
device-enforcement:Z-1185-2026
Official wording
Reason: due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.
Code information: Catalog/Model Numbers/ Lot Numbers / UDI codes: 10001 14250616-1019106 (01) 0 0850016 97600 0 (17) 251030 (10) SHOWCASE 10011 14250616-1019107 (01) 0 0850016 97609 3 (17) 251030 (10) SHOWCASE 10040c 50250624-1019120 (01) 0 0850016 97636 9 (17) 251030 (10) SHOWCASE 10040 50250624-1019119 (01) 0 0850016 97635 2 (17) 251030 (10) SHOWCASE 10032 14250616-1019108 (01) 0 0850016 97631 1 (17) 251030 (10) SHOWCASE 10041 50250522-1018643 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10041 50250624-1019121 (01) 0 0850016 97637 6 (17) 251030 (10) SHOWCASE 10021 15F250603-1018834 (01) 0 0850016 97619 2 (17) 251030 (10) SHOWCASE
Distribution pattern: U.S. Nationwide distribution in the states of AL, AR, AZ, CA, DE, FL, GA. IL, IN, KS, MI, MO, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN TX, UT, VA, and WA.
Derived failure modes
-
Unknown
due to a defect associated with the conductivity values of the solution used to calibrate meters and result in the reporting of incorrect conductivity values for dialysate.