Skip to content
Recall Observatory FDA recall evidence

Device product

Medline medical procedure convenience kits labeled as: KIT ROBOTICS UROLOGY PROSTATE, REF DYKMBNDL116G

Z-1090-2026

December 16, 2025

Class II

Product summary

Firm
Medline Industries, LP
Event
Event 98197
Status
Ongoing
Classification
Class II
Quantity
12 units
Official record key
device-enforcement:Z-1090-2026

Official wording

Reason: Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Code information: UDI/DI 10198459156991 (EA) 40198459156992 (CS), Lot Number 25EBM426

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.