Device product

BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Z-1008-2026

December 05, 2025

Class II

Product summary

Firm
Beaver-Visitec International, Inc.
Event
Event 98076
Status
Ongoing
Classification
Class II
Quantity
5 units
Official record key
device-enforcement:Z-1008-2026

Official wording

Reason: Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

Code information: Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.

Distribution pattern: US Nationwide distribution in the states of FL, MI, OK, TX, WI.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    Manufacturing deviation