Device product
Allura Xper FD10/10; Model Number: 722027;
Z-1074-2026
Product summary
- Event
- Event 98109
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 22 units
- Official record key
device-enforcement:Z-1074-2026
Official wording
Reason: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
Code information: Model Number: 722027; System Product Name: Allura Xper FD10/10; UDI-DI: 00884838054196; Serial Numbers: 298, 246, 146, 170, 238, 162, 7, 99, 81, 126, 55, 86, 6, 259, 205, 60, 57, 63, 145, 182, 48, 278;
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.
Derived failure modes
-
Unknown
Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.