Device product
RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System
Z-1102-2026
Product summary
- Event
- Event 98117
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 0
- Official record key
device-enforcement:Z-1102-2026
Official wording
Reason: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Code information: UDI: 0735000201031020201216. GTIN: 07350002010310. Serial Numbers: 10.1.0.613. Software Revision: RayStation 10B. Expiration Date: 2027-02-01.
Distribution pattern: Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Derived failure modes
-
Unknown
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.