Device product
CT 5300; Software Version Number: 4.5, 5.0, 5.1;
Z-0882-2026
Product summary
- Event
- Event 97919
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 6 units (OUS only)
- Official record key
device-enforcement:Z-0882-2026
Official wording
Reason: If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.
Code information: 1) Model Number: 728285; UDI-DI: 00884838113237; Serial Numbers: 35071, 800007, 800002, 800001, 800004, 800003;
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bhutan, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Ethiopia, France, FrenchPolynesia, Georgia, Germany, Greece, Guadeloupe, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Mexico, Moldova, Montenegro, Montserrat, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua Nw Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Rwanda, Saint Kitts, Saint Lucia, Saudi Arabia, Serbia, Sint Maarten (D, Slovakia, South Africa, South Korea, Spain, St.Pier,Miquel., Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Derived failure modes
-
Unknown
If the screws of tube heat exchanger on the rotating scanner are not tightened well after replacing. This component may become detached and make contact with other components located within the Incisive CT systems during rotation. Other components could be damaged due to the contact with the tube heat exchanger.