Device product
Extension Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog numbers: 1. EXTENSION SET, 18 IN; Catalog Number: 490448. 2. EXTENSION SET, 52 IN.; Catalog Number: 490465. 3. EXTENSION SET 49 IN.; Catalog Number: 490486. 4. EXTENSION SET, 12 CARESTITE LADS; Catalog Number: 490577.
Z-0630-2026
Product summary
- Event
- Event 97816
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 43,452 units
- Official record key
device-enforcement:Z-0630-2026
Official wording
Reason: Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Code information: 1. Catalog Number: 490448; Primary UDI-DI: 04046964848858; Unit of Dose UDI-DI: 04046964848841. Shelf Life: Earliest Exp of Comp or 36mths. 2. Catalog Number: 490465; Primary UDI-DI: 04046964950735; Unit of Dose UDI-DI: 04046964950728. Shelf Life: Earliest Exp of Comp or 36mths. 3. Catalog Number: 490486; Primary UDI-DI: 04046964991257; Unit of Dose UDI-DI: 04046964991240. Shelf Life: Earliest Exp of Comp or 24mths. 4. Catalog Number: 490577; Primary UDI-DI: 04046955287536; Unit of Dose UDI-DI: 04046955287529. Shelf Life: Earliest Exp of Comp or 36mths.
Distribution pattern: Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.
Derived failure modes
-
Unknown
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).