Device product
Elscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
Z-0291-2026
Product summary
- Event
- Event 97647
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 64 units
- Official record key
device-enforcement:Z-0291-2026
Official wording
Reason: GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Code information: all serial numbers in distribution
Distribution pattern: Worldwide
Derived failure modes
-
Unknown
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.