Skip to content
Recall Observatory FDA recall evidence

Device product

Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only

Z-0217-2026

September 12, 2025

Class II

Product summary

Firm
UIH Technologies LLC
Event
Event 97668
Status
Ongoing
Classification
Class II
Quantity
8 units
Official record key
device-enforcement:Z-0217-2026

Official wording

Reason: Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Code information: Serial Numbers: ; UDI: (01)06971576831012/ Serial Numbers: 300162 300163 300174 300176 304002 304003 304004 304006

Distribution pattern: US: OUS:

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.