Device product
CareLink Clinic, REF: MMT-7350
Z-0594-2026
Product summary
- Event
- Event 97873
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 12126
- Official record key
device-enforcement:Z-0594-2026
Official wording
Reason: Software error causing incorrect data to be displayed on the 24-hour Sensor Glucose Overview Graph, any potential therapy decisions were made based on the incorrect data displayed on the 24-hour sensor glucose overview graph may lead to hypoglycemia or hyperglycemia.
Code information: UDI-DI: 0763000B00008748K. Software Versions: 4.2B Software Available: US: Between September 4th, 2025, and October 3rd, 2025. OUS: Between September 11th, 2025, and October 4th, 2025. US (Clinical): Between August 21st, 2025, and October 3rd, 2025. OUS (Clinical): Between August 21st, 2025, and October 4th, 2025.
Distribution pattern: Worldwide Distribution: US (nationwide) including states of: TX, NC, MN, CA, VA, ID, WA, GA, MI, MA, OH, IL, IA, CT, FL, PA, NY, TN, AZ, LA, AL, NJ, CO, SC, MS, OR, MT, MD, KS, AK, NH, NE, WY, ND, IN, MO, WI, KY, UT, NV, SD, WV, NM, HI, AR, OK, VT, RI; and OUS (foreign) to countries of: Australia (AU), Sweden (SE), Belgium (BE), Great Britain (GB), Netherlands (NL), Finland (FI), Canada (CA), Germany (DE), Spain (ES), Switzerland (CH), Italy (IT), South Africa (ZA), South Korea (KR), Singapore (SG), New Zealand (NZ), Poland (PL), Israel (IL), Denmark (DK), Norway (NO), Luxembourg (LU), Lithuania (LT), Czech Republic ( CZ), Mexico (MX), Brazil (BR), Colombia (CO), Slovakia (SK), Ireland (IE), Argentina (AR), Greece (GR), Kuwait (KW), Portugal (PT), Iceland (IS), Japan (JP), Turkey (TR), Austria (AT), Qatar (QA), Hong Kong (HK), Chile (CL), Ukraine (UA), Hungary (HU), Estonia (EE), RS, Latvia (LV), Taiwan (TW), Saudia Arabia (SA), Bahrain (BH), Oman (OM), Romania (RO), India (IN), Uruguay (UY), Malaysia (MY), United Arab Emirates (AE), Algeria (DZ), Croatia (HR), France (FR), Russia (RU), Cyprus (CY), Thailand (TH), Egypt (EG), Bosnia and Herzegovina (BA), Libya (LY), and Iraq (IQ)
Derived failure modes
-
Device software or design
Software error causing incorrect