Skip to content
Recall Observatory FDA recall evidence

Device product

Cyclopen Microinterventional Cyclodialysis System. Model Number: CP-300. Product Description: The CycloPen Microinterventional Cyclodialysis device is a manual surgical instrument used for the ab-interno construction or modification of a cyclodialysis, followed by delivery of viscous materials, bio-tissue/allograft, or other materials to reinforce and maintain the cleft.

Z-0134-2026

September 06, 2025

Class II

Product summary

Firm
Iantrek Inc.
Event
Event 97640
Status
Ongoing
Classification
Class II
Quantity
1,262 units
Official record key
device-enforcement:Z-0134-2026

Official wording

Reason: Potential for compromised integrity of the sterile packaging.

Code information: 1. Model Number CP-300; UDI: +B962CP3000/$$7351645/16D20250201/14D20260201/Q1_; Lot Number: 351645; Exp date 02/2026. 2. Model Number CP-300; UDI: +B962CP3000/$$7351726/16D20250501/14D20260501/Q11; Lot Number 351726; Exp date 05/2026. 3. Model Number CP-300; UDI: +B962CP3000/$$7351795/16D20250601/14D20260601/Q19; Lot Number 351795; Exp date 06/2026.

Distribution pattern: US Nationwide distribution in the states of AL, AZ, CA, CT, DE, FL, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for compromised integrity of the sterile packaging.