Skip to content
Recall Observatory FDA recall evidence

Device product

The LeadCare¿ II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in whole blood. The LeadCare¿ II Test Kit is provided with the following materials: " Sensors (2 Containers of 24 sensors each; 48 tests) " Treatment Reagent Tubes (2 packages of 24 tubes each; 48 tests) " Heparinized Capillary Tubes/Plungers (50) " Transfer Droppers (50) " Calibration Button (1) " Lead Control Level 1 (1 @ 2 mL) " Lead Control Level 2 (1 @ 2 mL)

Z-0584-2026

February 23, 2024

Class II

Product summary

Firm
Magellan Diagnostics, Inc.
Event
Event 97858
Status
Ongoing
Classification
Class II
Quantity
1,774 units
Official record key
device-enforcement:Z-0584-2026

Official wording

Reason: Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.

Code information: Product Name: LeadCare¿ II Blood Lead Test Kit; Catalog Number: 70-6762; Impacted Lot/Sublot Numbers: 2333M, 2333M-01, 2333M-02, 2333M-03, 2333M-04, 2333M-05, 2333M-06, 2333M-07, 2333M-08, 2333M-09, 2333M-10, 2333M-11; Device Identifier: (01)00850355006000;

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Labeling or packaging

    reason.labeling_packaging · v1.0.0

    mislabeled