Device product
Basic Biopsy Tray, Item Number NMBP44L
Z-0170-2026
Product summary
- Event
- Event 97659
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 32,433 total
- Official record key
device-enforcement:Z-0170-2026
Official wording
Reason: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Code information: UDI-DI 191072188120 Lot 8515411 Entire lot impacted
Distribution pattern: US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Derived failure modes
-
Unknown
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.