Skip to content
Recall Observatory FDA recall evidence

Device product

ZAP-X Radiosurgery System, Model: 300150

Z-2639-2025

July 25, 2025

Class II

Product summary

Firm
Zap Surgical Systems
Event
Event 97413
Status
Ongoing
Classification
Class II
Quantity
17
Official record key
device-enforcement:Z-2639-2025

Official wording

Reason: If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

Code information: UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1

Distribution pattern: Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software defect