Device product
ZAP-X Radiosurgery System, Model: 300150
Z-2639-2025
Product summary
- Event
- Event 97413
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 17
- Official record key
device-enforcement:Z-2639-2025
Official wording
Reason: If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
Code information: UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1
Distribution pattern: Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.
Derived failure modes
-
Device software or design
software defect