Skip to content
Recall Observatory FDA recall evidence

Device product

Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) UIHY16AE ASC HYSTEROSCOPY PACK (PS 020634)

Z-0179-2026

September 04, 2025

Class II

Product summary

Firm
American Contract Systems Inc.
Event
Event 97659
Status
Ongoing
Classification
Class II
Quantity
32,433 total
Official record key
device-enforcement:Z-0179-2026

Official wording

Reason: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Code information: FHGY09S UDI-DI 191072234360 lot 8532011 bag serial number 68874879; LMGY39U UDI-DI 191072209337 Lot 8490211 bag serial numbers 47860009 47861175; OWGY11Q UDI-DI 191072236555 lot 8389711 bag serial number 17622382; RCGY52D UDI-DI 191072232670 lot 8142411 bag serial number 17487909; UIGN82M UDI-DI 191072106674 lot 8084811 bag serial numbers 17391081 17391091 17391111 17391171 lot 8445711 bag serial numbers 17683242 lot 8530511 bag serial numbers 17712404; UIHY16AE UDI-DI 00191072234117 Lot 8171811 N/A * If bag serial number is listed as "n/a", then entire lot is affected

Distribution pattern: US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.