Skip to content
Recall Observatory FDA recall evidence

Device product

Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40

Z-0131-2026

September 04, 2025

Class II

Product summary

Firm
Contego Medical LLC
Event
Event 97522
Status
Ongoing
Classification
Class II
Quantity
568
Official record key
device-enforcement:Z-0131-2026

Official wording

Reason: Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.

Code information: REF/UDI-DI/Lot(Expiration): NG-0740-140-2/00851616007613/Z2460707D(Z2460707D), Z2460733A(03/05/2027), Z2460766A(04/07/2027), Z2460774B(04/17/2027), Z2460780A(04/18/2027); NG-0640-140-2/00851616007590/Z2460746C(03/20/2027), Z2460774A(04/18/2027); NG-NV-7-40/00851616007620/Z2460776D(04/14/2027); NG-NV-6-40/00851616007651/Z2460776B(04/16/2027)

Distribution pattern: US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0