Device product
IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014
Z-0154-2026
Product summary
- Event
- Event 97698
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 23203
- Official record key
device-enforcement:Z-0154-2026
Official wording
Reason: IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
Code information: REF/UDI-DI/Lot: B33359/10887709007823/14070544, 14126963; B9213/10840619045859/14130197; B9732/10887709015033/14149636; B9733/10887709015040/14176779; 011-C7014/10840619005068/14029638, 14065820, 14110075, 14211300, 14129823
Distribution pattern: Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.
Derived failure modes
-
Unknown
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.