Device product
Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Z-0009-2026
Product summary
- Event
- Event 97502
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 87 units (21 US, 66 OUS)
- Official record key
device-enforcement:Z-0009-2026
Official wording
Reason: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Code information: Model Numbers: 722029; UDI-DI: 00884838054219; Serial Numbers: 71, 91, 7, 94, 62, 88, 36, 97, 21, 100, 30, 63, 60, 129, 95, 126, 110, 56, 55, 57, 31, 26, 49, 114, 128, 108, 103, 107, 81, 43, 52, 119, 118, 80, 78, 47, 106, 123, 96, 122, 14, 127, 101, 116, 102, 115, 19, 109, 120, 92, 69, 27, 46, 93, 51, 50, 124, 87, 20, 29, 32, 104, 34, 45, 59, 79, 24, 70, 39, 66, 82, 75, 40, 33, 23, 25, 85, 37, 44, 58, 38, 61, 98, 53, 89, 35, 48;
Distribution pattern: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Derived failure modes
-
Unknown
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.