Device product
Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DC use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.
Z-2538-2025
Product summary
- Firm
- Max Mobility LLC
- Event
- Event 97356
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 25,389 units
- Official record key
device-enforcement:Z-2538-2025
Official wording
Reason: Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.
Code information: All Serial Numbers distributed between the dates of April 25, 2022 - July 8, 2025
Distribution pattern: US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: CAN, NZL, KOR, BMU, GBR, JPN, TWN, AUS, BRA, COL, MEX, ABW, SWE, CHL, ARG, HKG
Derived failure modes
-
Unknown
Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.