Skip to content
Recall Observatory FDA recall evidence

Device product

Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH

Z-2612-2025

July 31, 2025

Class I

Product summary

Firm
Medline Industries, LP
Event
Event 97496
Status
Ongoing
Classification
Class I
Quantity
7913 eaches
Official record key
device-enforcement:Z-2612-2025

Official wording

Reason: Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

Code information: Item Number/UDI-DI 81102RH 10197344133239; 81104RH 10197344133246; 81107RH 10197344133260; 81174RH 10197344133291; 81402RH 10197344133352; 81404RH 10197344133376; 81405RH 10197344133383; 81472RH 10197344133413; 81473RH 10197344133420; 81474RH 10197344133437; 81531RH 10197344133512; 81532RH 10197344133529; LOTS EP250212 EP250214

Distribution pattern: Domestic US distribution nationwide. International distribution pending. No international distribution.

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    particulates