Device product
Medical convenience kits IHDC30R D AND C PACK
Z-0187-2026
Product summary
- Event
- Event 97659
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 32,433 total
- Official record key
device-enforcement:Z-0187-2026
Official wording
Reason: Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Code information: IHDC30R UDI-DI 191072226310 lot 8096511 bag serial number N/A * If bag serial number is listed as "N/A", then whole lot is affected
Distribution pattern: US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Derived failure modes
-
Unknown
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.