Device product
Salem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 266130 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) 266148 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 18 Fr/Ch 8888266106 Salem Sump" Dual Lumen Stomach Tube, Anti-Reflux Valve 10Fr/Ch (3.3 mm), 36" 266130CN Salem Sump" Tube 16FR with CONFIRM NOW, Nonsterile 266148CN Salem Sump" Tube 8 FR with CONFIRM NOW, Nonsterile
Z-0210-2026
Product summary
- Event
- Event 97539
- Status
- Ongoing
- Classification
- Class II
- Quantity
- all lots
- Official record key
device-enforcement:Z-0210-2026
Official wording
Reason: Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Code information: Product Code/Product Name/UDI-DI: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 10884521065024 266130 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) 10884521065031 266148 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 18 Fr/Ch 10884521065055 8888266106 Salem Sump" Dual Lumen Stomach Tube, Anti-Reflux Valve 10Fr/Ch (3.3 mm), 36" 20884521004655 266130CN Salem Sump" Tube 16FR with CONFIRM NOW, Nonsterile 10884521141292 266148CN Salem Sump" Tube 8 FR with CONFIRM NOW, Nonsterile 20884521141305 All Lots
Distribution pattern: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
Derived failure modes
-
Unknown
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.