Device product
RayStation with the following product descriptions: 1. RayStation 9B, Software Version: 9.1.0.933. 2. RayStation 9B Service Pack 1, Software Version: 9.2.0.483. 3. RayStation 10A, Software Version: 10.0.0.1154. 4. RayStation 10A Service Pack 1, Software Version: 10.0.1.52. 5. RayStation 10A Service Pack 2, Software Version: 10.0.2.10. Product Description: Radiation Therapy Treatment Planning System.
Z-2489-2025
Product summary
- Event
- Event 97400
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 26
- Official record key
device-enforcement:Z-2489-2025
Official wording
Reason: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Code information: 1. Software Version: 9.1.0.933, UDI: 0735000201026620191220. 2. Software Version: 9.2.0.483, UDI: 0735000201029720200310. 3. Software Version: 10.0.0.1154, UDI: 0735000201030320200526. 4. Software Version: 10.0.1.52, UDI: 0735000201036520200526. 5. Software Version: 10.0.2.10, UDI: 0735000201065520220608. Expiration Date: 2025-06-16
Distribution pattern: Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
Derived failure modes
-
Unknown
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.