Device product
Brand Name: ARVIS Product Name: ARVIS" Shoulder Model/Catalog Number: IN-27300, Arvis Surgical Planning Software Software Version: V2025.1.2 Product Description: Intended Use: The ARVIS" system is a computer-controlled navigation system intended to provide intra-operative measurements to the surgeon to aid in selection and positioning of orthopedic implant components. Description: ARVIS Shoulder is a computer-controlled surgical navigation system for shoulder arthroplasty. It aids the surgeon in making intra-operative measurements and locating anatomical structures of the shoulder joint based on the patient s preoperative imaging to assist with selection and positioning of orthopedic implant components. The system consists of software, electronic hardware and surgical instruments. The device s workflow involves CT based preoperative planning followed by intraoperative navigation and execution. The preoperative planning software enables 3D virtual implant positioning based on the patient s CT reconstructed digital bone model and bony landmarks. The shoulder navigation application software then matches the patient s digital bone model and landmarks to the intraoperative landmarks registered by the surgeon, allowing an image-based navigation to follow. The surgeon uses the plan data as guidance to navigate and help position shoulder instruments and implants.
Z-0145-2026
Product summary
- Event
- Event 97618
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 21 systems
- Official record key
device-enforcement:Z-0145-2026
Official wording
Reason: Complaint identified issue with AI surgical planning software that may result in implant malalignment and/or decrease range of motion.
Code information: Software Version V2025.1.2
Distribution pattern: US Nationwide distribution in the states of IN, MN, MO, OH, FL, IL, RI, KY, CA, WI, MI, PA, WA, CO, AL.
Derived failure modes
-
Device software or design
issue with AI surgical planning software