Skip to content
Recall Observatory FDA recall evidence

Device product

Ingenia 1.5T S Model Number (REF): 781347;

Z-2205-2025

June 30, 2025

Class II

Product summary

Firm
Philips North America
Event
Event 97172
Status
Ongoing
Classification
Class II
Quantity
136 units
Official record key
device-enforcement:Z-2205-2025

Official wording

Reason: The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

Code information: Model Number (REF): 781347; UDI-DI: 00884838068421; Serial Numbers: 81001, 81003, 81004, 81005, 81006, 81007, 81008, 81009, 81010, 81011, 81013, 81014, 81015, 81016, 81017, 81018, 81019, 81020, 81023, 81024, 81025, 81027, 81028, 81029, 81030, 81031, 81032, 81033, 81034, 81035, 81037, 81038, 81040, 81041, 81042, 81043, 81044, 81045, 81046, 81047, 81048, 81049, 81050, 81051, 81052, 81055, 81056, 81058, 81059, 81060, 81061, 81062, 81063, 81064, 81065, 81067, 81070, 81071, 81072, 81074, 81075, 81076, 81077, 81078, 81079, 81080, 81084, 81085, 81086, 81087, 81088, 81090, 81091, 81092, 81093, 81094, 81095, 81096, 81097, 81098, 81103, 81105, 81106, 81108, 81111, 81112, 81113, 81114, 81115, 81116, 81117, 81118, 81122, 81123, 81124, 81126, 81127, 81128, 81129, 81131, 81132, 81133, 81137, 81138, 81139, 81140, 81145, 81146, 81147, 81148, 81149, 81151, 81152, 81153, 81154, 81155, 81156, 81157, 81158, 81159, 81162, 81164, 81165, 81167, 81170, 81171, 81172, 81173, 81174, 81175, 81176, 81178, 81179, 81180, 81184, 81185;

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, French Guiana, FrenchPolynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen, Zambia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.