RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Z-2230-2025

June 26, 2025

Class II

Product summary

Firm
Siemens Healthcare Diagnostics Inc
Event
Event 97209
Status
Ongoing
Classification
Class II
Quantity
1,477 units
Official record key
device-enforcement:Z-2230-2025

Official wording

Reason: Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

Code information: Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.

Distribution pattern: worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    manufacturing error