Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.
Z-2420-2025
Product summary
- Firm
- MAQUET GMBH
- Event
- Event 97313
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 315 units (8 US, 307 OUS)
- Official record key
device-enforcement:Z-2420-2025
Official wording
Reason: Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
Code information: Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; UDI-DI: (1) 04046768145177, (2) 04046768145184, (3) 04046768145191, (4) 04046768145207, (5) 04046768145214, (6) 04046768145221; Lot/Serial Numbers: (1) 1, 2, 3, 10, 25, 26, 27, 28, 22, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 24, (2) 9, 51, 53, 14, 15, 16, 17, 34, 35, 36, 37, 39, 40, 42, 43, 54, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 38, 44, 45, 46, 5, 13, 57, 58, 59, 18, 49, 47, 48, 7, 52, (3) 41, 5, 6, 45, 48, 49, 50, 23, 38, 10, 12, 13, 14, 84, 87, 88, 89, 7, 63, 80, 82, 83, 86, 54, 93, 94, 11, 15, 18, 20, 21, 22, 27, 28, 32, 40, 9, 69, 70, 71, 72, 73, 74, 75, 76, 77, 78, 79, 85, 55, 56, 57, 58, 59, 42, 97, 92, 17, 26, 16, 24, 25, 30, 33, 35, 36, 37, 39, 44, 53, 91, 8, 19, 90, 46, 47, 51, 52, 34, 43, 60, 62, 64, 65, 66, 67, 68, (4) 44, 70, 71, 42, 56, 65, 115, 185, 52, 105, 106, 107, 109, 145, 146, 147, 148, 11, 26, 19, 32, 33, 35, 36, 39, 53, 61, 63, 69, 14, 20, 68, 13, 18, 16, 101, 167, 170, 83, 3, 4, 23, 28, 37, 38, 40, 41, 43, 45, 47, 48, 50, 55, 58, 60, 62, 78, 81, 82, 92, 94, 96, 97, 98, 99, 100, 102, 116, 138, 139, 144, 150, 152, 153, 154, 155, 156, 165, 169, 171, 21, 22, 24, 25, 6, 17, 27, 29, 30, 34, 49, 51, 57, 64, 74, 75, 76, 77, 79, 80, 84, 85, 86, 87, 88, 89, 90, 91, 93, 95, 103, 104, 108, 110, 111, 112, 113, 114, 118, 119, 120, 121, 122, 123, 124, 125, 126, 127, 128, 129, 130, 131, 132, 133, 134, 135, 136, 137, 140, 141, 142, 143, 172, 173, 174, 194, (5) 1, 2, 3, 4, 5, 6, 7, (6) 12, 15;
Distribution pattern: Worldwide - US Nationwide distribution in the states of KY, NY, TX and the countries of Algeria, Australia, Austria, Bulgaria, Canada, China, Colombia, Denmark, Egypt, Estonia, Finland, France, Germany, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kuwait, Lebanon, Morocco, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom.
Derived failure modes
-
Unknown
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.