Skip to content
Recall Observatory FDA recall evidence

Device product

Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014

Z-2519-2025

July 31, 2025

Class II

Product summary

Firm
Lumicell, Inc.
Event
Event 97316
Status
Ongoing
Classification
Class II
Quantity
200 units (20 packs containing 10 units per pack)
Official record key
device-enforcement:Z-2519-2025

Official wording

Reason: The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.

Code information: Model/Catalog Number: 900-00218 UDI-DIs: Sterile Cover 10-Pack Shipper Box Assembly: (01)0 0860008 53781 2, 840-00204 Assy, Dispenser Carton Sterile Cover: (01)0 0860003 25251 2, 810-00210 Assy, Outer Tray Sterile Cover: (01) 0 0860003 2520 5, Lot numbers: 43930, 43931, 43932, 43933, 42756, 44349, 44349-01, 44349-02, 44349-03, 44349-04;

Distribution pattern: US Nationwide distribution in the states of CA, FL, MA, NC, TX.

Derived failure modes

  • Foreign material or chemical contamination

    reason.foreign_material_chemical_contamination · v1.0.0

    foreign bodies