Device product

Inseego USB8 4G Dongle Kit, Part Number: 01-2298, which contains USB Flash Drive, Part Number: 43-2011. The kit is an accessory to the Tempus Pro Monitor, REF: 00-1007-R, 00-1024-R, 00-1026-R

Z-2175-2025

May 30, 2025

Class II

Product summary

Firm
Remote Diagnostic Technologies Ltd.
Event
Event 97197
Status
Ongoing
Classification
Class II
Quantity
55
Official record key
device-enforcement:Z-2175-2025

Official wording

Reason: Kit USB flash drive contains outdated software, which could result in reverting patient monitor to outdated software and reintroducing the video laryngoscope issue (related to an August 2023 recall), which could lead to delay in diagnosis, delayed treatment, hypoxia due to unexpected loss of video laryngoscopy and all other monitor measurements during system restart.

Code information: REF/UDI-DI: 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932. Software Version: vx.28

Distribution pattern: Worldwide - US Nationwide distribution in the states of SC, KY, NV, IA, MI and the country of GB.

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software and reintroducing the video laryngoscope issue