Device product
Servo-n Ventilator System. Model Number: 6694800.
Z-2014-2025
Product summary
- Event
- Event 97032
- Status
- Ongoing
- Classification
- Class I
- Quantity
- 7,688 units
- Official record key
device-enforcement:Z-2014-2025
Official wording
Reason: Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
Code information: Model Number: 6694800. UDI-DI: 07325710001110. All software versions.
Distribution pattern: US Nationwide distribution and Puerto Rico. No international distribution.
Derived failure modes
-
Unknown
Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.